The SPIROVIT SP-1 is a unique solution to simplify complex lung function testing. The high resolution user-friendly screen displays real-time flow/volume loops and detailed measurement matrices. The data can be either printed within seconds or transmitted to a PC.
Features:
- FVC, SVC, MVV, MV and pre/post medication tests with choice of predicted normal
- Simple operation
- Internal memory for up to 100 tests (optional)
- Built-in mains and battery operation
- Ready-to-file reports within seconds on 3.5” (90 mm) paper
- Comparison of predicted and actual values
- Automatic interpretation
- Data transmission via built-in RS-232 interface to PC-based data management program SEMA for long term storage, retrieval and trending
- Everything in one box … memory, battery, printer, patient incentive screen
- High quality accessories for maximum precision, e.g. absolute humidity insensitive (SCHILLER sensors) SP-150
Includes:
- 1 pack of recording paper
- 10 flow tubes
- SP-150 sensor (DTC)
- 1 noseclip
- 1 pack of disposable mouthpieces (for SP-150)
- 1 power cable
- 1 operating manual
Specifications:
- Dimensions - 290 x 210 x 69 mm, app. 2.9 kg; 11.4 x 8.3 x 2.7 in, app. 6.3 lbs
- Power Supply Requirements:
- 220–240V, nominal 230V, 50/60 Hz, or 100–115V, nominal 115V, 50/60 Hz
- Mains independent operation with built-in rechargeable battery and charger unit, LED indication for line or battery operation
- Battery Charging Time - 15 hours from a completely discharged battery (< 3 hours to 60 % capacity)
- Battery Capacity - 6 hours of normal use
- Power Consumption - Max. 28VA
- Storage - Memory for up to 100 tests
Liquid Crystal Display:
- Liquid crystal display for graphic and alphanumeric representation
- Resolution - 192 x 64 dots
- Contrast is manually variable via keyboard
Chart Paper - Thermoreactive, Z-folded, width 90 mm/3.6in, perforation each page
Printing Process - High-resolution thermal printhead, 8 dots/mm
Paper Speed - 25 mm/s
Safety Standards - MDD 93/42/EEC, Annex II; EN 60601-1: 1990; IEC 60601-1-2
Safety Class - BF according to IEC 60601-1
Protection Class:
- I according to IEC 60601-1, CSA 22.2 and UL 2601 (units with internal battery)
- Ila according to MDD 93/42/EEC (medical guideline)
Conformity: -0123 according 93/42/EEC
Control Panel and Keyboard - User-friendly Rubber Keys, LED indicators
Environmental Conditions:
- Temperature, operating: 10° to 40° C/50° to 104° F
- Temperature, storage: -10° to 50° C/14° to 122° F
- Relative Humidity - 25 to 95 % (not condensed)
- Pressure, operating - 700–1060 hPa
Measured Values:
FVC: FVC, FEV0.5, FEV1.0, FEV3.0, FEV0.5/FVC, FEV1.0/FVC, FEV3.0/FVC, FEF0.2–1.2, FEF25–75%, FEF75–85%, PEF, MEF75%, MEF50%, MEF25%, FIVC, FIV1.0, FIV1.0/FIVC, FIV1.0/FVC, PIF, MIF50%
SVC: SVC, ERV, IRV, TV
MV: MV, RR, TV
MVV: MVV, RR, TV
Prediction Equation:
Adults - ECCS, Crapo, Morris, Knudson, Knudson 76, Austria, Berglund, Finnish, Indian, Composite, Polgar
Children - Quanier&Tammeling, Austria, India, Knudson, Knudson 76, Polgar
Comparison pre/post medication possible